Isoflurane concentrations using uncompensated vaporisers within circle systems.
Abstract
In a series of studies designed to investigate the need for a temperature-compensated vaporiser for use in a circle system, we first studied temperature changes within and isoflurane concentration delivered by a Komesaroff vaporiser during bench tests using different gas flows. Agent temperature and vapour concentration decreased as predicted by the Antoine equation. Using the vaporiser within a circle system during clinical anaesthesia, we then studied 20 patients breathing spontaneously and a further 10 patients receiving controlled ventilation, measuring the temperature of the agent within the vaporiser and the concentration of agent inspired by a the patient. In clincial use with the frest gas flows of 1-3.min-1, the inspired agent concentration did not decrease despite the decrease in temperature of the liquid isoflurane in the vaporiser.