Abstract
Conventional 3D dose calculations for stereotactic radiosurgery involve integration of individual static beams comprising a set of arcs. For iterative optimization of multiple isocentre treatment, which requires repetitive dose calculations at a large number of sample points, the conventional method is too slow. To overcome this problem spherically symmetric dose distributions are assumed. The authors describe a spherical dose model derived from a parametrized convolution of the collimator width and a dose spread kernel. The method is fast and easy to implement requiring just a single empirically derived value. Furthermore, the model is differentiable with respect to the parameters to be optimized. This property is useful when the optimization strategies rely on gradient information.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
P S Cho, H G Kuterdem, R J Marks. 1998. A spherical dose model for radiosurgery plan optimization.. https://doi.org/10.1088/0031-9155%2F43%2F10%2F032