PubMed · 9817903
Multicenter evaluation of a modified protocol for the RAPIDEC staph system for direct identification of Staphylococcus aureus in blood cultures.
Abstract
A modified protocol for the RAPIDEC Staph system (bioMérieux, Marcy-l'Etoile, France) for direct identification of Staphylococcus aureus in blood cultures was evaluated in a multicenter study. A total of 129 blood cultures (BACTEC 9000 Blood Culture System; Becton Dickinson Diagnostic Instrument Systems, Sparks, Md.) containing gram-positive cocci in clusters were analyzed by conventional methods and by RAPIDEC Staph in accordance with the manufacturer's protocol and in accordance with a modified protocol. The sensitivity, specificity, and positive and negative predictive values obtained with the manufacturer's protocol were 90.5, 97.7, 95. 0, and 95.5%, respectively, and those obtained with the modified protocol were 100, 96.6, 93.3, and 100%, respectively. The modified protocol for the RAPIDEC Staph is easier to perform than the manufacturer's protocol and is very reliable.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
A van Griethuysen, A Buiting, W Goessens, P van Keulen, R Wintermans, J Kluytmans. 1998. Multicenter evaluation of a modified protocol for the RAPIDEC staph system for direct identification of Staphylococcus aureus in blood cultures.. https://doi.org/10.1128/jcm.36.12.3707-3709.1998
Cite the original work for its findings. Save a collection to share your selection of sources.