PubMed · 10501073
Establishing a quality control program for an automated dosimetry system.
Abstract
Automatic dosimetry systems can provide instantaneous dose and dose-rate information during fluoroscopic procedures as well as long-term records of patient doses. For this information to be useful, it is necessary that the accuracy of such systems be maintained through a rigorous quality control program. Daily and weekly quality control checks were performed on a PEMNET automated dosimetry system to determine its stability and the value of such tests in a quality control program. Weekly tests included monitoring the accuracy of the measured doses under a variety of conditions. The results of the tests indicate possible improvements in test methodology and real and potential sources of system failure and provide a statistical basis for setting quality control limits for future system monitoring.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
S Li, T J O'Dea, R A Geise. 1999. Establishing a quality control program for an automated dosimetry system.. https://doi.org/10.1118/1.598665