A quantitative investigation of additive noise reduction for active matrix flat-panel imagers using compensation lines.

Abstract

A quantitative investigation of a technique for reducing correlated noise in indirect detection active matrix flat-panel imagers has been reported. Correlated noise in such systems arises from the coupling of electronic noise, originating from fluctuations in external sources such as power supplies and ambient electromagnetic sources, to the imaging array via its address lines. The noise reduction technique involves the use of signals from columns of compensation line pixels located in relatively close proximity to the columns of normal imaging pixels on the array. Compensation line pixels are designed to be as sensitive to externally-coupled noise as columns of normal imaging pixels but are insensitive to incident radiation. For each imaging pixel, correlated noise is removed by subtracting from the imaging pixel signal a signal derived from compensation line pixels located on the same row. The effectiveness of various implementations of this correction has been examined through measurements of signal and noise from individual pixels as well as of noise power spectra. These measurements were performed both in the absence of radiation as well as with x rays. The effectiveness of the correction was also demonstrated qualitatively by means of an image of a hand phantom. It was found that the use of a single compensation line dramatically reduces external noise through removal of the correlated noise component. While this form of the correction increases non-radiation-related uncorrelated noise, the effect can be largely reduced through the introduction of multiple compensation lines. Finally, a position-dependent correction based on compensation lines on both sides of the array was found to be effective when the magnitude of the correlated noise varied linearly across the array.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Y El-Mohri, L E Antonuk, Q Zhao, M Maolinbay, X Rong, K W Jee, S Nassif, C Cionca. 2000. A quantitative investigation of additive noise reduction for active matrix flat-panel imagers using compensation lines.. https://doi.org/10.1118/1.1287053

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Compliance of electronic personal neutron dosemeters with the new International Standard IEC 61526.

The recommendations and test requests for the dose equivalent response of personal neutron dosemeters formulated by the new International Standard IEC 61526 are summarised. In particular, IEC 61526 allows the use of broad fields if dosemeters do not fulfil the hard requirements using monoenergetic neutrons. Some broad fields which can work as a replacement field using ISO sources (252Cf, 252Cf (D2O mod.), 241Am-Be) and simulated workplace fields (CANEL and SIGMA) are described. This work shows the results of recent measurements of the personal dose equivalent response for the dosemeters Thermo Electron EPD-N2, Aloka PDM-313 and the prototype dosemeter PTB DOS-2002, and discusses their compliance with respect to the new IEC 61526 standard.

Electronics↗

Treatment of waste printed wire boards in electronic waste for safe disposal.

The printed wire boards (PWBs) in electronic waste (E-waste) have been found to contain large amounts of toxic substances. Studies have concluded that the waste PWBs are hazardous wastes because they fails the toxicity characteristic leaching procedure (TCLP) test with high level of lead (Pb) leaching out. In this study, two treatment methods - high-pressure compaction and cement solidification - were explored for rendering the PWBs into non-hazardous forms so that they may be safely disposed or used. The high-pressure compaction method could turn the PWBs into high-density compacts with significant volume reduction, but the impact resistance of the compacts was too low to keep them intact in the environment for a long run. In contrast, the cement solidification could turn the PWBs into strong monoliths with high impact resistance and relatively high compressive strength. The leaching of the toxic heavy metal Pb from the solidified samples was evaluated by both a dynamic leaching test and the TCLP test. The dynamic leaching results revealed that Pb could be effectively confined in the solidified products under very harsh environmental conditions. The TCLP test results showed that the leaching level of Pb was far below the regulatory level of 5mg/L, suggesting that the solidified PWBs are no longer hazardous. It was concluded that the cement solidification is an effective way to render the waste PWBs into environmentally benign forms so that they can be disposed of as ordinary solid wastes or beneficially used in the place of concrete in some applications.

Electronics↗

Evaluation of dose equivalent by the electronic personal dosemeter for neutron 'Saphydose-N' at different workplaces of nuclear facilities.

This paper presents the results of measurements made with the electronic personal neutron Saphydose-N during the four campaigns of the European contract EVIDOS (EValuation of Individual DOSimetry in mixed neutron and photon radiation fields). These measurements were performed at Institute for Radiological Protection and Nuclear Safety (IRSN) in France (C0), at the Krümmel Nuclear Power Plant in Germany (C1), at the VENUS Research Reactor and the Belgonucléaire fuel processing plant in Belgium (C2) and at the Ringhals Nuclear Power Plant in Sweden (C3). The results for Saphydose-N are compared with reference values for dose equivalent.

Electronics↗