Impedance analyser module for EIT and spectroscopy using undersampling.

Abstract

In this paper we present the concept, the design and the test procedure for a DSP-based high-precision and high-performance wide-band (up to 10 MHz) bioimpedance analyser module for application in EIT or bioimpedance spectroscopy. The module implements a digital concept with appropriate signal conditioning hardware for voltage and current measurement, early signal digitization and subsequent digital signal processing in order to calculate the components of impedance (or admittance). At low frequencies, the module utilizes the conventional direct conversion method, whereas at high frequencies the undersampling technique is used. The advantages of the described system are the following: (a) the frequency range is extended to higher frequencies, (b) the number of data sampled per time interval is significantly reduced, and (c) the current consumption and the costs of the ADCs can be significantly reduced. The validation procedure is performed by comparing the measured and theoretical values of the magnitude and the phase of the impedance for a commonly used tissue model. The module offers an accuracy of better than 0.012% for the magnitude of impedance and better than 0.02 degrees for the phase.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

T Dudykevych, E Gersing, F Thiel, G Hellige. 2001. Impedance analyser module for EIT and spectroscopy using undersampling.. https://doi.org/10.1088/0967-3334%2F22%2F1%2F303

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The effect of ohmic heating on vacuum drying rate of sweet potato tissue.

Ohmically heating fruit and vegetable tissue has been shown to increase hot-air drying rate, shift desorption isotherms, and increase juice extraction yields with respect to untreated, conventionally heated, and microwaved samples. The objective of this study was to determine if ohmically heating sweet potato tissue would enhance the vacuum drying rate of these samples with respect to untreated samples. Sweet potato cubes were ohmically heated to three endpoint temperatures using three electrical field strengths and were then placed in a freeze dryer. Moisture content vs. time data were collected and modeled. Results showed that the vacuum drying rates of ohmically heated samples were faster than raw samples for most treatment combinations, and that the maximum reduction of drying time was 24%. Minimal ohmic treatment can result in a significant decrease in vacuum drying time, which could have important economic and product quality implications.

Electric Impedance↗

Disruption of nasopharyngeal epithelium by pneumococci is density-linked.

BACKGROUND: The aim of this project was to study the influence of pneumococci on nasopharyngeal epithelial integrity as a function of time and pneumococcal density. MATERIALS AND METHODS: Cell layers of an in vitro model of human nasopharyngeal epithelium were inoculated with different pneumococcal strains. The transepithelial electrical resistance (TEER), a measure of the integrity of the cell layers, and the pneumococcal concentration in the apical fluid on the epithelial cells were measured at different times after inoculation. RESULTS: Pneumococci caused a decrease in the TEER when a density of 1 x 107 CFU mL-1 was reached. The growth rate of pneumococci in our in vitro model differed between the strains tested and, for the same strain, between in vitro culture on the epithelial cells and broth culture. Differences in timing of the onset of decrease in the TEER between strains were the result of differences in growth rate on the epithelial cells. Antibiotic-induced lysis of pneumococci caused an immediate decrease in the TEER of the cell layers. CONCLUSION: Pneumococci cause a decrease in the TEER at a density of 1 x 107 CFU mL-1. Our hypothesis is that this decrease in the TEER is the result of quorum-induced lysis of the pneumococci.

Electric Impedance↗

Batch electrophoretic cells with Eyring fluids: analysis of the hydrodynamics.

Mixing and dispersion phenomena caused by the carrier fluid in an electrophoretic cell is the main subject of this study. In particular, the effects of Joule heating on temperature and velocity profiles for Eyring-model fluids (EMF) are studied. The heat transfer is sequentially coupled with momentum transfer to derive an analytical expression for both the temperature and the velocity profiles. These results are then used to show the hydrodynamic behavior of the fluid in a batch electrophoretic cell. Furthermore, the results obtained are useful to compare with the fluid behavior of other carriers of different rheology, such as Newtonian fluids, power-law fluids, and viscoelastic fluids that obey the CEF model. The results show that EMF are potentially good carriers for relatively high Joule heat generation and therefore good candidates to control mixing inside the electrophoretic cell.

Electric Impedance↗