Energetic parameter changes with mechanical ventilation in conjunction with BVAD assistance.

Abstract

The aim was to assess the influence of a biventricular assist device (BVAD) on ventricular energetics parameters (external work, oxygen consumption, cardiac mechanical efficiency) for both ventricles, when mechanical ventilation was applied. The experiments were performed using a computer simulator of cardiovascular system (CARDIOSIM) after modelling a pathological state of the left ventricle (E(v)Left = 0. 9 mmHg cm(-3) and increasing pulmonary resistance (Rap = 0.3 mmHg cm(-3 s). The effect of mechanical ventilation was mean intrathoracic pressure changes from 0 to +5 mmHg. This simulation showed that application of BVAD for both ventricles reduces external work and that this effect is stressed by positive intrathoracic pressure, reduces cardiac mechanical efficiency that is quite insensitive to intrathoracic pressure and increases oxygen consumption, which is reduced by positive intrathoracic pressure. The increase of potential energy at the onset of BVAD evidences a rightwards shift of ventricular work cycle (unloading of the ventricles). In general, positive intrathoracic pressure during BVAD assistance adversely affects ventricular energetics.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

C De Lazzari, M Darowski, G Ferrari, F Clemente, M Guaragno. Energetic parameter changes with mechanical ventilation in conjunction with BVAD assistance.. https://doi.org/10.1080/03091900210127924

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Optimal design of non-Newtonian, micro-scale viscous pumps for biomedical devices.

The present paper addresses the numerical optimization of geometrical parameters of non-Newtonian micro-scale viscous pumps for biomedical devices. The objective is to maximize the mass flow rate per unit of shaft power consumed by the rotor when an external pressure load is applied along the channel that houses the rotor. Two geometric parameters are considered in the optimization process: (i) the height of the channel that houses the rotor (H) and (ii), the eccentricity (epsilon) of the rotor. Three different micro-scale viscous pump configurations were tested: a straight-housed pump (I-shaped housing) and two curved housed pumps (L- and U-shaped housings). The stress-strain constitutive law is modeled by a power-law relation. The results show that the geometric optimization of micro-scale viscous pumps is critical since the mass flow rate propelled by the rotor is highly dependent on epsilon and H. Numerical simulations indicate that mass flow rate is maximized when epsilon approximately 0, namely when the rotor is placed at a distance of 0.05 radii from the lower wall. The results also show that micro-scale viscous pumps with curved housing provide higher mass flow rate per unit of shaft power consumed when compared with straight-housed pumps. The results are presented in terms optimized dimensions of all three configurations (i.e., H(opt) and epsilon(opt)) and for values of the power-law index varying between 0.5 (shear thinning fluids) and 1.5 (shear-thickening fluids).

Biomedical Engineering↗