PubMed · 12135458
Sampling effects influence heights measured with atomic force microscopy.
Abstract
The atomic force microscope (AFM) is an exquisitely delicate probe measuring the height of a specimen at discrete sampling points in a fixed two-dimensional (2D) raster. The resulting topograph is a 2D digital image, with each pixel representing a distinct height measurement. The height of an object is determined as the average of the maximum heights measured above the supporting surface. We show that such object heights derived from a variety of organic samples depend critically on the sampling or pixel size of the 2D raster. It is concluded that to obtain accurate specimen heights, the pixel size must be small enough to resolve submolecular structures and thus ensure representative sampling of the height variation on the surface.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J B Heymann, C Möller, D J Müller. 2002. Sampling effects influence heights measured with atomic force microscopy.. https://doi.org/10.1046/j.1365-2818.2002.01039.x