PubMed · 1258508
[Arrhythmia analysis--arrhythmia control (comparative studies of 4 small computers for automatic electrocardiography control)].
Abstract
Four arrhythmia computers presently available on the market were tested with respect to their efficiency in identifying abnormal QRS-complexes. A considerable variation for the calculated sensitivity and specificity has been found with a highest value for sensitivity (92.9%) combined with a low specificity (72.4%), and a highest value for specificity (100%) combined with a very low sensitivity (36.6%). The best results were obtained by a computer with a sensitivity of 90.9% and a specificity of 98.5%. For the routine use of these computers in the CCU common artifacts (e.g. muscle potentials) must influence their performance in a practical way: neither should minor artifacts stop the function completely nor should the computer continue to measure with a high level of "noise". Based upon the results of these test series suggestions for the concept of arrhythmia computers for CCU are presented.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
P Kühn, A Kroiss, G Joskowicz. 1976. [Arrhythmia analysis--arrhythmia control (comparative studies of 4 small computers for automatic electrocardiography control)].. https://pubmed.health/record/pubmed%3A1258508/arrhythmia-analysis-arrhythmia-control-comparative-studies-of-4-small-computers-for-automatic-electr