PubMed · 15496732
CCD-based X-ray area detector for time-resolved diffraction experiments.
Abstract
A fast X-ray area detector for diffraction, scattering and imaging experiments at microsecond to millisecond time resolution has been developed. The key element of the detector is a fast (291 frames s(-1)) framing camera with three CCDs. A prism forms identical images on the CCDs and the frame rate is increased three times by reading them alternately. In order to convert X-rays into visible light that is detectable with the CCDs, an X-ray image intensifier is used. The camera can also be used with a high-resolution X-ray detector. In both cases it was found to be important to use a phosphor with a short decay time to fully make use of the high-speed framing capability of the camera. Preliminary results of a fibre diffraction experiment on a skeletal muscle and coronary angiography are presented.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Naoto Yagi, Katsuaki Inoue, Toshihiko Oka. 2004-10-22. CCD-based X-ray area detector for time-resolved diffraction experiments.. https://doi.org/10.1107/s090904950402093x
Cite the original work for its findings. Save a collection to share your selection of sources.