Peak shape improvement of basic analytes in capillary liquid chromatography.
Abstract
The analysis of bases is of special interest in pharmaceutical research because numerous active substances contain basic functional groups. Capillary and conventional size LC separations of drug substances spiked with potential impurities were compared. In the case of the nonpolar drug levonorgestrel equal separation efficiency was readily attained by both techniques. The peaks of basic substances, however, showed extensive tailing when separated by capillary LC. The peak deformation was attributable to interactions of the basic substances with the polar inner surface of the fused silica capillaries employed in capillary LC and does not appear with the steel tubing generally used in conventional size LC. This drawback of capillary LC was overcome by use of deactivated fused silica capillaries for column hardware and transfer lines.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Anja Prüss, Christine Kempter, Jens Gysler, Jürgen Maier-Rosenkranz, Thomas Jira. 2005. Peak shape improvement of basic analytes in capillary liquid chromatography.. https://doi.org/10.1002/jssc.200401807