PubMed Health⌕ Search

PubMed · 16652179

A method for dynamic system characterization using hydraulic series resistance.

Abstract

The pressure required to drive flow through a microfluidic device is an important characteristic of that device. We present a method to measure the flow rate through microfluidic components and systems, including micropumps and microvalves. The measurement platform is composed of two pressure sensors and a glass tube, which provides series resistance. The principle of the measurement is the fluid dynamical equivalent of Ohm's law, which defines the relationship between current, resistance, and voltage that are analogues to flow rate, hydraulic resistance, and pressure drop, respectively. Once the series resistance is known, it is possible to compute the flow rate through a device based on pressure alone. In addition, the dynamic system characteristics of the device-resistance and capacitance-can be computed. The benefits of this method are its simple configuration, capability of measuring flow rate accurately from the more easily measured pressure, and the ability to predict the dynamic response of microfluidic devices.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Dongshin Kim, Naomi C Chesler, David J Beebe. 2006-03-20. A method for dynamic system characterization using hydraulic series resistance.. https://doi.org/10.1039/b517054k

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Nonlithographic fabrication of microfluidic devices.

A facile nonlithographic method for expedient fabrication of microfluidic devices of poly(dimethylsiloxane) is described. Positive-relief masters for the molds are directly printed on smooth substrates. For the formation of connecting channels and chambers inside the polymer components of the microfluidic devices, cavity-forming elements are adhered to the surfaces of the masters. Using this nonlithographic approach, we fabricated microfluidic devices for detection of bacterial spores on the basis of enhancement of the emission of terbium (III) ions.

Microfluidic Analytical Techniques↗

Reactions in droplets in microfluidic channels.

Fundamental and applied research in chemistry and biology benefits from opportunities provided by droplet-based microfluidic systems. These systems enable the miniaturization of reactions by compartmentalizing reactions in droplets of femoliter to microliter volumes. Compartmentalization in droplets provides rapid mixing of reagents, control of the timing of reactions on timescales from milliseconds to months, control of interfacial properties, and the ability to synthesize and transport solid reagents and products. Droplet-based microfluidics can help to enhance and accelerate chemical and biochemical screening, protein crystallization, enzymatic kinetics, and assays. Moreover, the control provided by droplets in microfluidic devices can lead to new scientific methods and insights.

Microfluidic Analytical Techniques↗

Periodic microfluidic bubbling oscillator: insight into the stability of two-phase microflows.

This Letter describes a periodically oscillating foam flow in a microfluidic device. For constant input parameters, both the produced bubble volume and the flow rate vary over a factor of two. We make explicit the link between the foam topology alternance and flow rate changes, and construct a retroaction model where bubbles still present downstream determine the volume of new bubbles, in agreement with experiment. This gives insight into the various parameters important to maintain monodispersity and at the same time shows a method to achieve controlled polydispersity.

Microfluidic Analytical Techniques↗