PubMed Health⌕ Search

PubMed · 16711661

Acute microelectrode array implantation into human neocortex: preliminary technique and histological considerations.

Abstract

OBJECT: Researchers at The Center for Neural Interfaces at the University of Utah have designed and produced a silicon-based high-density microelectrode array that has been used successfully in mammalian models. The authors investigate the ability to transfer array insertion techniques to humans and examine the acute response of human cortical tissue to array implantation. METHODS: Six patients who were scheduled to undergo temporal lobectomy surgery were enrolled in an Institutional Review Board-approved protocol. Before the patients underwent lateral temporal cortical resection, one or two high-density microelectrode arrays were implanted in each individual by using a pneumatic insertion device. Cortical tissue was then excised and preserved in formalin. The specimens were sectioned and stained for histological examination. Pneumatic insertion of a microelectrode array into human cortex in the operating room was feasible. There were no clinical complications associated with implantation and no evidence of significant insertion-related hemorrhage. Tissue responses ranged from mild cortical deformity to small focal hemorrhages several millimeters below the electrode tines. Based on initial results, the insertion device was modified. A footplate that mechanically isolates a small area of cortex and a calibrated micromanipulator were added to improve the reproducibility of insertion. CONCLUSIONS: A high-density microelectrode array designed to function as a direct cortical interface device can be implanted into human cortical tissue without acute clinical complications. Further modifications to the insertion device and array design are ongoing and future work will assess the functional significance of the tissue reactions observed.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Paul A House, Joel D MacDonald, Patrick A Tresco, Richard A Normann. 2006-05-15. Acute microelectrode array implantation into human neocortex: preliminary technique and histological considerations.. https://doi.org/10.3171/foc.2006.20.5.5

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The development of a multichannel electrode array for retinal prostheses.

The development of a multielectrode array is the key issue for retinal prostheses. We developed a 10 x 10 platinum electrode array that consists of an 8-microm polyimide layer sandwiched between 5-microm polymonochloro-para-xylylene (parylene-C) layers. Each electrode was formed as a 30-microm-high bump by Pt/Au double-layer electroplating. We estimated the charge delivery capability (CDC) of the electrode by measuring the CDCs of two-channel electrode arrays. The dimensions of each electrode of the two-channel array were the same as those of each electrode formed on the 10 x 10 array. The results suggest that for cathodic-first (CF) pulses, 80% of electrodes surpassed our development target of 318 microC/cm2, which corresponds to the charge density of pulses of 500 micros duration and 200 microA amplitude for a 200-microm-diameter planar electrode.

Electrodes, Implanted↗

Improved version of the printed circuit board (PCB) modular multi-channel microdrive for extracellular electrophysiological recordings.

The modular multi-channel PCB microdrive was described some years ago, since then several improvements were introduced while using these drives. Utilizing several years of experience with the original PCB microdrive we redesigned it to improve its stability and usability. The application of the printed circuit board technology and the extensive use of flexible fused silica capillaries for fabrication of the microdrive are described in detail. The improved design led to a low cost and light-weight multi-channel microdrive with outstanding modularity for extracellular field, single unit or multiunit tetrode recording up to 64/128 channels.

Electrodes, Implanted↗