Programming eccentricities of an activity sensing pacemaker.

Abstract

The Activitrax rate responsive pacemaker system has enjoyed wide popularity but minor engineering eccentricities have occurred and have been reported. We report another unusual feature seen in Activitrax models 8400, 8402, and 8403. This feature consists of continuing in a temporary mode while a different permanent mode was programmed. These eccentricities of the programming features are not very commonly seen, however, they can be somewhat perplexing to the physician following the patient. These unusual features are no longer present in the newer models being manufactured.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R B Leman, R D McVenes, J M Kratz, P C Gillette. 1990. Programming eccentricities of an activity sensing pacemaker.. https://doi.org/10.1111/j.1540-8159.1990.tb01995.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

FFT-based digital tactile vocoder system for real-time use.

A microprocessor-based real-time digital vibrotactile vocoder system has been developed to train the deaf and for artificial hearing research. The system is composed of a microcomputer module with a digital signal processor, interface units and an attenuator/driver circuit. Live or digitised (stored or synthetic) speech is presented to the skin spectrally through a belt housing eight or 16 vibrators. Speech is processed in real time using a fast Fourier transform. The system is also capable of presenting any arbitrary spatiotemporal pattern on the skin for artificial hearing experiments. A preliminary experiment with a deaf subject indicates that the system is potentially an effective device for artificial hearing.

Biomedical Engineering