PubMed · 17201852
Optimization of device programming for cardiac resynchronization therapy.
Abstract
Cardiac resynchronization therapy may lead to remarkable improvement in clinical status in selected patients with heart failure. However, approximately 20-30% of patients may not respond to this treatment. One of the reasons for this may be suboptimal programming of the device, which has particular considerations as compared to standard pacemakers. Hemodynamic response to pacing may be affected by timing of the atrioventricular (AV) interval, affecting synchronicity of atrial and ventricular contraction. In addition current biventricular devices have separate right and left ventricular channels that allow programming of an interventricular (VV) interval with right or left ventricular preexcitation. This article focuses on the parameters that may be optimized for biventricular pacing, and reviews the different techniques currently available for this application, with special emphasis paid to echocardiography.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Haran Burri, Henri Sunthorn, Dipen Shah, René Lerch. 2006. Optimization of device programming for cardiac resynchronization therapy.. https://doi.org/10.1111/j.1540-8159.2006.00557.x
Cite the original work for its findings. Save a collection to share your selection of sources.