PubMed · 2614303
Simultaneous bipolar pacing and sensing during EPs.
Abstract
Bipolar pacing and sensing-recording on the same electrode is a problem during pacemaker implantation as well as during electrophysiological studies. A method is described that makes this possible. By using bipolar simultaneous sensing-pacing on the same electrode, the system can record local repolarization after a depolarization controlled by stimulation. The bipolar-paced evoked potential is useful as an indicator to drive a bipolar QT rate responsive pacemaker. Even more, during EPS it can be used to detect capture during tachycardia stimulation and to measure the exact refractory period.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J Snoeck, M Berkhof, C Vrints. 1989. Simultaneous bipolar pacing and sensing during EPs.. https://doi.org/10.1016/s0022-0736(07)80129-4
Cite the original work for its findings. Save a collection to share your selection of sources.