A paediatric ventilator with a fluidic control system.
Abstract
A paediatric ventilator has been designed basically as a fluidic-logic controlled T-piece occluder. Inspiratory and expiratory time, inflation pressure limit, positive end-expiratory pressure (PEEP), continuous positive airway pressure (CPAP) and inspiratory flow-rate are controlled independently and intermittent mandatory ventilation (IMV) is available. Warning systems are provided for failure of the driving as, low airway pressure and inspiratory pressure limit. The breathing circuit is isolated from the control and warning systems for ease of sterilization.