PubMed HealthSearch

PubMed · 3936457

Cementless acetabular replacement using a pegged polyethylene prosthesis.

Abstract

Acetabular replacement without cement using three types of pegged polyethylene prostheses was begun as a prospective trial by Morscher (August 1977). Ring (June 1979), and Freeman (January 1980); 1878 hips have been replaced using this technique and are reviewed. Follow-up averaged 2 years (range: 6 months-6 years). Successful pain relief and satisfactory walking ability were present in 97% and adequate flexion in 90%. Radiographic review showed the consistent development of a stable, uniform sclerotic line adjacent to the prosthesis. No component migration could be documented. Ten hips have been revised for obvious femoral loosening, and two acetabula showed slight motion during revision for femoral failure and were revised. Deep infection (0.37%) occurred only in patients not receiving perioperative antibiotics and not operated upon in unidirectional filtered air-flow enclosures. Postmortem histological evaluation of eight hips showed a thin fibrous membrane separating the prosthesis from viable bone. These results are preliminary, but nevertheless at least as good as those in any large series of cemented acetabula with the same follow-up. The advantages of this technique are its reproducible results, low complication rate, limited invasion of the skeleton, and easy revision should failure occur.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

K C Bertin, M A Freeman, E Morscher, A Oeri, P A Ring. 1985. Cementless acetabular replacement using a pegged polyethylene prosthesis.. https://doi.org/10.1007/bf00450220

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Migration of porous coated acetabular prostheses fixed with screws: roentgen stereophotogrammetric analysis.

The fixation of screw-fixed, porous-coated acetabular prostheses was studied during 2 years in 21 patients (22 hips) using repeated roentgen stereophotogrammetric measurements. Migration of the center of the prostheses and/or tilting were recorded in five cups after 6 months. Two years after operation, migration was recorded in three, migration and rotation in six, and only rotation in three cups. Bone quality, prosthetic position, and immediate postoperative radiographic appearance of the prosthesis-bone interface did not correlate with the occurrence of migration. Narrow radiolucent zones had developed in half of the patients after 2 years and was associated with proximal migration. Micromovements of screw-fixed, press-fit cups seem to start later compared with previously presented results from cemented acetabular prosthesis.

Acetabulum

Polygonal triple osteotomy of the pelvis. A correction for dysplastic hip joints.

A new triple osteotomy, polygonal pelvic osteotomy (PPO), is described for use in dysplastic hips. The hip joint must be congruous with a centre-edge (CE) angle between +20 degrees and -10 degrees. Templates are used before and during operation which allow precise correction of the position of the acetabulum. The geometry of the osteotomy and a special compression plate achieve sufficient stability to permit early mobilisation. Surgery is indicated in adolescents and young adults if pain persists for at least two months. The surgical exposure, operative technique and postoperative management are presented, with the two-year results in nine cases.

Acetabulum