A system oriented electrocardiographic amplifier.
Abstract
A low noise, programmable gain amplifier has been developed for acquiring ECG data from a variety of recording environments. Features include: low noise (4muV peak-to-peak) differential inputs; digitally programmable gains (range 100 to 16000) and output DC offsets (range -3.5 to +4 V) controlled via a serial interface; externally selectable low frequency response (0.04 or 0.08 Hz); and an on-board output monitor multiplexer controlled via a second serial interface.