PubMed · 524640
A wearable artificial glomerulus.
Abstract
We have shown that hemofiltration can be conducted for 16 hrs/day for more than 3 continuous days by connecting an Amicon 0.2 M2 hemofilter to an arteriovenous shunt. We can produce over 8 ml/min of ultrafiltrate with no external pump, and this ultrafiltrate can be replaced by having the patient drink an electrolyte-glucose solution. Anticoagulation can be attained with heparin, warfarin and cyproheptadine. Each filter can be expected to function at least 48 hrs before it must be replaced.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
M S Neff, S Sadjadi, R Slifkin. 1979. A wearable artificial glomerulus.. https://doi.org/10.1097/00002480-197902500-00014