PubMed · 6596837

A signal processor for a multiple-electrode hearing prosthesis.

Abstract

A 22-electrode implantable hearing prosthesis uses a wearable speech processor which estimates three speech signal parameters. These are voice pitch, second formant frequency and flattened spectrum amplitude. The signal is monitored continuously for periodicity in the range 80-400 Hz and, if this is present, stimulation occurs at the same rate. Otherwise, as in the case of unvoiced sounds, it occurs at the random rate of fluctuation of the signal envelope. The second formant is obtained by filtering to extract the dominant peak in the midband region and by continuous measurement of the zero crossing rate. The amplitude measured is that of the whole speech spectrum, pre-emphasized by differentiation. The values that are presented to the patient are the parameter estimates immediately prior to the stimulation pulse. Second formant frequency is coded by selection of an appropriate electrode in the cochlea and amplitude by a suitably controlled current. Automatic gain control is used to keep the dynamic range of the amplitude estimate within the 30 dB range of the circuitry.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

P M Seligman, J F Patrick, Y C Tong, G M Clark, R C Dowell, P A Crosby. 1984. A signal processor for a multiple-electrode hearing prosthesis.. https://pubmed.ncbi.nlm.nih.gov/6596837/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The restoration of hearing in neurofibromatosis type 2.

Central electrical stimulation of the auditory pathway can allow hearing in patients suffering from deafness localized in the auditory nerve. Developments in a multi-channel auditory brainstem implant based on the Nucleus Mini 22 Cochlear implant with transcutaneous signal transmission is discussed. The devices have been implanted in nine European patients suffering from Neurofibromatosis Type 2. Preliminary speech perception results and patient satisfaction are encouraging, and the data presented include some limited open speech recognition.

Cochlear Implants

An unusual cause of cochlear implant failure.

We present a cochlear implant failure previously unidentified and unreported. Following successful implantation and activation of a Cochlear Mini 22 system, a distressing barrage of crackling noises was heard by the patient despite repeated programming with threshold and comfort level adjustment. The implant soon became nonfunctional and integrity testing showed deficient electrical responses and abnormal surface measured wave forms. Investigation of the explant, following reimplantation with an identical system, showed a defective trim capacitor, an electric component of the internal circuit in the stimulator. Paired Student's t-test showed a marginally significant increase (p = 0.047) in effective threshold level and a highly significant decrease (p = 0.0002) in maximum comfort level for the second implant compared to the first implant.

Cochlear Implants