PubMed · 6682983
Fluoroscopic dose reduction using a digital television noise-reduction device.
Abstract
A digital video image processor, connected to a video system in a conventional pediatric fluoroscopy room, was used to determine whether the device could provide satisfactory fluoroscopic images during routine examinations when the x-ray tube was operated at substantially lower than normal radiation-dose levels. A 50% reduction resulted in image quality which was indistinguishable from conventional fluoroscopic views.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R C Ablow, C C Jaffe, S C Orphanoudakis, R I Markowitz, N S Rosenfield. 1983. Fluoroscopic dose reduction using a digital television noise-reduction device.. https://doi.org/10.1148/radiology.148.1.6682983. https://pubmed.health/record/pubmed%3A6682983/fluoroscopic-dose-reduction-using-a-digital-television-noise-reduction-device
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite PubMed Health by Indiciae: BibTeX · RIS.