Correction for beam hardening in computed tomography.
Abstract
Corrections for beam-hardening artifacts in computed tomography can be made by using a model which assumes that water and bone mineral are the only constituents of tissue. With this model, a correction factor for the measured transmission values can be calculated such that the reconstructed attenuation coefficients have values corresponding to a monoenergetic source of known energy. Systematic errors in the uncorrected attenuation coefficients, which may be 5%, can be reduced to less than 1% if corrected transmission values are used.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
P K Kijewski, B E Bjärngard. Correction for beam hardening in computed tomography.. https://doi.org/10.1118/1.594429