PubMed · 6802916
A flexible system for closed-loop ventilator development.
Abstract
The feasibility of using a microcomputer to regulate the delivery of artificial ventilation was investigated experimentally. To carry out this study, a sampled-data proportional, integral, derivative feedback controller was implemented on a commercial microcomputer and interfaced with a digital/analogue model of a prototype ventilator and two-compartment patient. The microcomputer sampled end-tidal CO2 and calculated a new frequency of ventilation from the difference of end-tidal CO2 and a desired set point value. Use of a microcomputer programmable in BASIC allowed easy manipulation of feedback structure. With the proper choice of feedback constants, the microcomputer was able to maintain end-tidal CO2 at a desired set point.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
W I Marsh, W D Smith. 1982. A flexible system for closed-loop ventilator development.. https://doi.org/10.1007/bf00994120
Cite the original work for its findings. Save a collection to share your selection of sources.