An animal model for testing automatic defibrillators.
Abstract
A promising therapy for ventricular fibrillation--a life-threatening cardiac arrhythmia--is implantation of an automatic defibrillator. A critical component of such a device is the system that detects the presence of ventricular fibrillation. Automatic systems for detecting ventricular fibrillation have been tested with arrhythmias produced by electric shocks in normal canine hearts, but have not been tested with spontaneous arrhythmias in hyperirritable hearts. We have developed an animal model to create arrhythmias without electrical stimulation and have used it to test our automatic defibrillator. This model permits evaluation of both reliability to diagnose VF and reliability to reject other tachyarrhythmias.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J D Bourland, C F Babbs, W A Tacker, L A Geddes. An animal model for testing automatic defibrillators.. https://pubmed.ncbi.nlm.nih.gov/7354730/