PubMed · 9605459
A computer based, temperature controlled bipolar electrocoagulation system: II. Optimal temperature and coagulation time.
Abstract
OBJECTIVE: A computer based, temperature controlled bipolar coagulation system has been developed. This system was evaluated for its efficiency on electrocoagulation of rabbit uterine horns at a low temperature. STUDY DESIGN: In purpose to find an optimal setting, rabbit uterine horns were coagulated at various temperature and time. They were examined histologically and histochemically up to six weeks after coagulation. RESULTS: Electrocoagulation at 90 degrees C for 30 s induced a total occlusion of the lumen when examined six weeks later. CONCLUSION: This coagulation system with low temperature is sufficient for a total occlusion of uterine lumen in the rabbit.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
K Kato, M Reich, K Bühler, A E Schindler. 1998. A computer based, temperature controlled bipolar electrocoagulation system: II. Optimal temperature and coagulation time.. https://doi.org/10.1016/s0301-2115(98)00017-7