Solid-state fluoroscopic imager for high-resolution angiography: parallel-cascaded linear systems analysis.

Abstract

Cascaded linear systems based modeling techniques have been used in the past to predict important system parameters that have a direct impact on image quality. Such models are also useful in optimizing system parameters to improve image quality. In this work, detailed analysis of a solid-state fluoroscopic imaging system intended for high-resolution angiography is presented with the use of such a model. The imaging system analyzed through this model uses four 8 x 8 cm three-side buttable interlined charge-coupled devices (CCDs) specifically designed for high-resolution angiography and tiled in a seamless fashion to achieve a field of view (FOV) of 16 x 16 cm. Larger FOVs can be achieved by tiling more CCDs in a similar manner. The system employs a CsI:Tl scintillator coupled to the CCDs by straight (nontapering) fiberoptics and can potentially be operated in 78, 156, or 234 microm pixel pitch modes. The system parameters analyzed through this model include presampling modulation transfer function, noise power spectrum, and detective quantum efficiency (DQE). The results of the simulations performed indicate that DQE(0) in excess of 0.6 is achievable, with the imager operating at 156 microm pixel pitch, 30 frames/s, and employing a 450-microm-thick CsI:Tl scintillator, even at a low fluoroscopic exposure rate of 1 microR/frame. Further, at a nominal fluoroscopic exposure rate of 2.5 microR/frame there was no noticeable degradation of the DQE even at the 78 microm pixel pitch mode suggesting that it is feasible to perform high-resolution angiography hitherto unattainable in clinical practice.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Srinivasan Vedantham, Andrew Karellas, Sankararaman Suryanarayanan. 2004. Solid-state fluoroscopic imager for high-resolution angiography: parallel-cascaded linear systems analysis.. https://doi.org/10.1118/1.1689014

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Transhepatic vascular access for diagnostic and interventional procedures: techniques, outcome, and complications.

Early or multiple cardiac catheterizations may result in occlusion of the femoral veins prohibiting their use. The internal jugular or sub-clavian approach may be an appropriate alternative. However, these approaches may not be suitable in patients with surgical interruption of the superior vena cava. In other patients, they may not allow easy access to certain areas of the heart. The transhepatic approach is an important alternative route for performing cardiac catheterization and interventions. Depending on the planned procedure, it may be the preferred route to perform the cardiac catheterization even in the presence of patent femoral veins. The indications, technical details and potential complications of this approach are discussed.

Angiography↗

Embolization for a bleeding pelvic fracture in a patient with persistent sciatic artery.

Emergency embolization for a bleeding pelvic fracture was performed in a patient with persistent sciatic artery, a rare anatomic variation in which the internal iliac artery continues to the popliteal artery. Successful hemostasis was achieved without complications by means of selective catheterization into each branch of the internal iliac artery. Awareness of this unusual vascular anatomy is critically important to avoid serious complication of emergency embolization.