Defects in GT rotary instruments after use: an SEM study.

Abstract

Used rotary nickel-titanium instruments require frequent replacement. This laboratory study evaluated defects of GT Rotary instruments before and after usage. The instruments were fixed into custom-made holders and photographed in an SEM at x 260 to x 12,000 magnification at preset points. Instrument sizes 20 with 1.0, 0.8, and 0.6 taper and instrument sizes 35 with 1.2 taper were used in 12 canals. The used instruments were cleaned and then reexamined in an SEM as before. The presence of 11 types of conditions was scored from the pre- and postusage photographs. No instruments fracturing or plastic deformation was observed during the test. The presence of debris, pitting, and scraping changed significantly. Imperfections were found on used GT Rotary instruments and the presence of defects (pitting, scraping, and blunt-cutting edges) increased in all of the instruments. The behavior of GT Rotary instruments with usage seems to be different, compared with that found in other instruments.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

T R Tripi, A Bonaccorso, V Tripi, G G Condorelli, E Rapisarda. 2001. Defects in GT rotary instruments after use: an SEM study.. https://doi.org/10.1097/00004770-200112000-00018

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The effect of wear on ultrasonic scaler tip displacement amplitude.

AIM: During clinical usage, scaler tips may become worn and reduced in length. It is unknown what effect wear has on the magnitude of scaler tip vibrations when they are utilized under typical clinical loads. The aim of this investigation was to assess the effect of simulated wear on ultrasonic scaler tip displacement amplitude, using a scanning laser vibrometer. MATERIALS AND METHODS: A Cavitron SPS (Dentsply) ultrasonic generator and three scaler insert designs (FSI-100, FSI-1000 and FSI-SLI-10S) were selected for the investigation. Tip vibration displacement amplitude was assessed unloaded and then contacting against tooth surfaces with loads of 0.5 and 1.0 N. Tips were then ground down by 1 mm and then 2 mm and scans were repeated. RESULTS: For all tips, load and length were found to be significant variables (p<0.0001). The scaler tips showed a fall in displacement amplitude with a reduction in tip length. However, all scaler tips showed variability in the amount oscillation that occurred. This was most pronounced with FSI-SLI-10S. CONCLUSIONS: This investigation demonstrated that tip wear could affect the performance of dental ultrasonic scaler inserts by reducing their vibration displacement amplitude. Clinicians should be aware of this variability, which may be significant enough to affect clinical procedures.

Dental High-Speed Technique↗

What's the rush?

Explore the source record for details and available documents.

Dental High-Speed Technique↗